Quality Assurance Low Contrast Test (Optional)
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.5-9
0.0
Performing the low contrast test
(for other countries)
F.5
Preparing the phantom
F.5
F.5
(1) Water phantom
(2) Slice thickness phantom
(3) Wire phantom
(4) Low contrast phantom
✧ Attach the low-contrast phantom section to the Siemens
phantom combination, using the mounting bracket and 4
mounting screws.
✧ Mount the phantom assembly on the patient table at nor-
mal body height.
✧ Move the patient table into the gantry until the scan center
is aligned with the center of the low contrast phantom.