syngo CT 2007E Emotion Operator Manual
02/2008
0.0
Constancy Test of the Monitor Quality Assurance
F.7-6
The individual tests
F.7
Performing the luminance
measurements
F.7
In this test the veiling luminance (environmental lighting) and
the maximum contrast are tested.
F.7
Switch the monitor off and wait at least 30 seconds.
Adjust then the room lighting to the operating conditions
using the dimmer switch.
Measure veiling luminance Ls at the center of the screen
with the luminance meter.
i
.7
F
When taking measurements with the luminance meter, observe
the distance recommended by the manufacturer (40 - 60 cm).
F.7
Switch on the monitor.
Select test image 1 to measure the minimum and maxi-
mum luminance.
i
.7
F
In test image 1, the squares for the maximum greyscale value
(L
max.
, bright field) and the minimum grayscale value (L
min.
,
black field) are used.
F.7