Quality Assurance Constancy Test of the Monitor
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.7-7
0.0
With the luminance meter, measure the minimum contrast
(L
min.
) and the maximum contrast (L
max.
).
The ratio L
max.
/L
min.
is calculated as maximum contrast MK
( 40:1).
F.7
i
.7
F
You can simply increase the contrast ratio by reducing the envi-
ronmental lighting intensity. However, make sure that enough
light is available for reading and writing.
F.7
Testing the spatial and contrast
resolution
F.7
The visual resolution is checked in this test.
F.7
Select test image 2 or test image 3.
Check the lines of the grid with a modulation of 100% high
contrast (HC) at the center and in the four corners.
The lines of the grid must be recognizable.