Safety Software Related Safety Information
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
A.4-33
0.0
syngo Neuro Perfusion CT
A.4
syngo Neuro Perfusion CT provides images of blood flow, blood
volume and time to peak from one set of dynamic CT images.
A.4
Image prerequisites
A.4
syngo Neuro Perfusion CT was designed and tested for the
processing of dynamic image sequences of the brain on
SOMATOM CT systems.
A.4
A.4
A.4
CAUTION
A.4
Image requirements for the intended perfusion calculation not
fulfilled!
A.4
Misleading interpretation of results.
A. 4
Take special care when processing dynamic CT images
other than SOMATOM CT image data.
If processing other levels of the brain, ensure that the slice
contains the occipital segment of the superior sagittal sinus
above the confluence of sinuses.
Only use data that have been acquired or reconstructed
with a sampling rate of at least 1 image per second.
Note that the primary anatomy that the software expects
for its semi-automatic analysis are standard slices through
the basal ganglia.