Exceptional situations
D. 1
Special circumstances can force the system to shut itself down
or to reduce its function temporarily until a defined operating
state is achieved.
D. 1
There are three types of exceptional situations:
D. 1
❏ Overheating or overcooling of the gantry
❏ Overheating of the computer system
❏ Power failure
The system responds to such exceptional situations in several
steps:
D. 1
Countdown
D. 1
If the system reaches a critical state, a countdown is started.
During this time, do not perform any examinations and termi-
nate any current application(s).
D. 1
Recovery
D. 1
If the system returns to the normal state during countdown
(temperature dropped, power supply restored), you can resume
examination.
D. 1
Shutdown
D. 1
If the countdown has elapsed, the system or the components
affected are shut down.
D. 1
Restart
D. 1
If the system or a component is shut down, a restart is per-
formed once the exceptional situation has been remedied.