syngo CT 2007E Emotion Operator Manual
02/2008
0.0
Information about Personal Safety Safety
A.2-10
Special instructions on HeartView CT
A.2
A.2
A.2
Pacemaker patients
A. 2
Rate maters may continue to count the pacemaker rate during
cardiac arrest or arrhythmias.
A.2
Keep pacemaker patients under close surveillance.
A.2
CAUTION
A.2
Using a combination of different electronic devices on one
patient!
A.2
The total leakage current may exceed safety limits.
A. 2
Do not use more electronic devices on the patient than
absolutely necessary.
A.2
CAUTION
A.2
Conductive parts become live!
A.2
Heart arrhythmia.
A. 2
Ensure patient safety and make sure the conductive parts of
the ECG electrodes (including associated connectors) and
other patient-applied parts do not contact other conductive
parts, including earth ground, at any time.