syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.7-1
0.0
CHAPTER
F.6
F.7
Constancy Test of the
Monitor
Monitors are subjected to a constancy test to ensure that the
image quality and measured values lie within a defined range.
F.7
Constancy test requirements for image display devices (moni-
tors) may vary according to different national regulations.
F.7
The constancy test described herein complies with the German
X-ray regulations (Deutsche Roentgenverordnung) and is based
on requirements given in DIN V 6868-57.
F.7
i
.7
F
We recommend that the constancy test and collection of the
reference values is performed and documented by authorized
personnel.
F.7
Image quality of the monitor
F.7
The acceptance test performed during initial start up commis-
sioning ensures optimum image quality of the device. The
results of these measurements are documented in an accept-
ance test report and serve as reference values for the follow up
constancy test.
F.7
B
.7
F
If any alterations are made to the device that affect the image
quality, e.g., repairs, replacement of parts, or readjustments,
the acceptance test according to the Germany X-ray regulations
must be repeated.
F.7