Software Related Safety Information Safety
A.4-64
Storage commitment
A.4
The work status of a data object indicates the processing stage
of patient and examination data. This status is indicated as an
abbreviation in every list entry in the content area.
A.4
A.4
A.4
CAUTION
A.4
The flags A (Archive) and S (Send) respectively only indicate the
receipt of the images. They do not indicate successful storage in
the intended archive!
A.4
T h e d a t a m a y b e l o s t i f i t i s d e l e t e d b y t h e s e n d e r e . g . , b y a n a u t o
delete mechanism and if it cannot be stored by the receiver.
A. 4
Before deleting data from your local database ensure that
this data is safely stored in the intended archive. Always use
'storage commitment' if supported by sender and receiver
of data.