Safety Software Related Safety Information
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
A.4-63
0.0
A.4
CAUTION
A.4
The system blocks when the audit trail is filled (too many
records, or too much drive space covered)!
A.4
In this case, it is not possible to work with the system. Emergency
access is also not possible.
A. 4
To prevent system blockage, someone must archive the log
files and remove them from the audit trail on a regular
basis.
When the audit trail blocks the system, proceed as
described in operator manual.
Note that then the "SecuritySystem > BypassBlockedSys-
tem" and the "AuditTrail>Archive" privileges are needed.