syngo CT 2007E Emotion Operator Manual
02/2008
0.0
Use of Accessories System Handling
C.4-4
Using the scanning footswitch
C.4
In interventional CT examinations, a scan is always controlled
with the footswitch in the examination room.
C.4
For radiation protection please refer to (→ Page A.2-22 Radia-
tion protection).
C.4
You can use the scanning footswitch and the buttons on the
gantry control panel in parallel.
C.4
C.4
Triggering the scan
C.4
✧ Press the footswitch and keep it pressed.
The scan is performed for as long as the footswitch remains
pressed.
C.4
Terminating the scan
C.4
✧ Release the footswitch.
Scanning is terminated. The system is again ready to scan.
C.4
C.4
CAUTION
C.4
Use of unsuitable footswitches!
C.4
Incorrect function possible.
C.4
✧ Only an original Siemens footswitch must be installed.