Safety Information about Personal Safety
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
A.2-35
0.0
System related artifacts
A.2
System related artifacts may be due to the following:
A.2
Uncalibrated system
Detector deviations
A.2
A.2
CAUTION
A.2
When you start or restart the system, the detector has not yet
reached operating temperature!
A.2
Incorrect diagnosis resulting from image artifacts.
A. 2
Calibrate the system as part of the checkup. Repeat calibra-
tion after about 60 minutes (Setup > Calibration), or earlier
if ring artifacts occur.