System related artifacts
A.2
System related artifacts may be due to the following:
A.2
❏ Uncalibrated system
❏ Detector deviations
A.2
A.2
CAUTION
A.2
When you start or restart the system, the detector has not yet
reached operating temperature!
A.2
Incorrect diagnosis resulting from image artifacts.
A. 2
✧ Calibrate the system as part of the checkup. Repeat calibra-
tion after about 60 minutes (Setup > Calibration), or earlier
if ring artifacts occur.