EasyManua.ls Logo

Siemens SOMATOM Emotion syngo CT 2007E - Page 78

Siemens SOMATOM Emotion syngo CT 2007E
574 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
Information about Personal Safety Safety
A.2-34
Metal artifacts
A. 2
Metal objects such as tooth fillings, surgical clips, jewelry, hair
clips, belts or hip prostheses etc. can cause extreme forms of
beam hardening artifacts. Depending on density and size, they
can lead to total absorption of the radiation. This results in cor-
responding strong black or white streaks or star-shaped arti-
facts.
A.2
Motion artifacts
A. 2
Movement of organs in the slice or displacement of the entire
slice during scanning likewise causes bright and dark artifacts.
These are usually seen as streaks or in areas of low density. To
reduce motion artifacts use the automatic patient instructions
(API) function for breathing and swallowing commands.
A.2
Greater difficulties are encountered with peristalsis. Heartbeat
artifacts are also problematic.
A.2
In sequence mode, a special reconstruction technique, the
Motion Artifact Correction Algorithm (MCA) is used to further
reduce motion artifacts (not selectable).
A.2

Table of Contents

Related product manuals