Switching the System On and Off Operation
D.1-8
System checkup and calibration
D. 1
After syngo has started up, you should start system checkup
and calibration to enable the system for scanning.
D. 1
The checkup procedure is started in the Checkup dialog box
which will be displayed after system startup. If the Checkup dia-
log box does not appear automatically, select Setup > Checkup.
D. 1
i
D.1
If the last checkup has been completed no longer than 12 hours
ago, the time since the last complete checkup is indicated.
D. 1
You have the option of skipping the checkup. If necessary, you
can calibrate the system later (Setup > Calibration).
D. 1
D. 1
D. 1
CAUTION
D. 1
When you start or restart the system, the detector has not yet
reached operating temperature!
D. 1
Incorrect diagnosis resulting from image artifacts.
D. 1
✧ Calibrate the system as part of the checkup. Repeat calibra-
tion after about 60 minutes (Setup > Calibration), or earlier
if ring artifacts occur.