0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
Constancy Test of the Monitor Quality Assurance
F.7-2
Measuring and test equipment
F.7
The constancy test is performed using the following test equip-
ment:
F.7
❏ Luminance meter for LCD monitors
Specification: Class B (DIN 5032-7) with valid calibration,
range 0.05 cd/m
2
– 1000cd/m
2
(recommended device:
Spotmeter SMfit Act for LCD monitors)
F.7
❏ Test image source (imaging device)
❏ Test images:
Technical images (1,2,3)
F.7
Clinical reference images (4,5,6)
F.7