Safety Information about Personal Safety
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
A.2-19
0.0
A.2
CAUTION
A.2
Uncontrolled system movements and radiation!
A.2
Injury or radiation damage.
A. 2
✧ Press STOP if a system movement is performed that has not
been initiated on a gantry operator panel, on the control
box or during a scan (esp. at autorange).
✧ Press STOP if the patient table moves in the wrong direc-
tion.
✧ Always observe the patient during system movements.
✧ Press STOP if a key sticks or a movement does not stop
immediately when the key is released.
✧ Press STOP if the patient table does not stop as expected.
✧ Press STOP if the patient table position displayed differs by
more than 10 mm from the actual position.
✧ Press STOP if the patient is not correctly positioned during
system movements.
✧ Press STOP if the HOLD key does not respond during a scan.