Safety Information about Personal Safety
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
A.2-19
0.0
A.2
CAUTION
A.2
Uncontrolled system movements and radiation!
A.2
Injury or radiation damage.
A. 2
Press STOP if a system movement is performed that has not
been initiated on a gantry operator panel, on the control
box or during a scan (esp. at autorange).
Press STOP if the patient table moves in the wrong direc-
tion.
Always observe the patient during system movements.
Press STOP if a key sticks or a movement does not stop
immediately when the key is released.
Press STOP if the patient table does not stop as expected.
Press STOP if the patient table position displayed differs by
more than 10 mm from the actual position.
Press STOP if the patient is not correctly positioned during
system movements.
Press STOP if the HOLD key does not respond during a scan.