syngo CT 2007E Emotion Operator Manual
02/2008
0.0
Switching the System On and Off Operation
D.1-16
Switching off the system :
D. 1
If no errors occur during shutdown, the computer will be
switched off.
D. 1
Switch off the UPS: press the OFF pushbutton at the UPS for
at least 3 seconds.
Switch off the gantry power switch at the LCB.
The scan system will be switched off.
D. 1
B
D.1
Unless absolutely necessary, do not switch off the system at the
main switch provided by the customer or the EMERGENCY OFF
switch. Otherwise you will have to calibrate the system several
times when you restart it. (→ Page D.1-8 System checkup and
calibration)
D. 1