syngo CT 2007E Emotion Operator Manual
02/2008
0.0
Switching the System On and Off Operation
D.1-28
Restarting the system
D. 1
✓ Normal power supply is reactivated
D. 1
✧ Restart the system. (→ Page D.1-1 Switching the System
On and Off)
D. 1
CAUTION
D. 1
The detector heating is switched off as long as the system is dis-
connected from the power supply. When you restart the system,
the detector has not yet reached operating temperature!
D. 1
Incorrect diagnosis resulting from image artifacts.
D. 1
✧ Calibrate the system as part of the checkup. Repeat calibra-
tion after about 60 minutes (Setup > Calibration), or earlier
if ring artifacts occur.