0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
Cleaning Agents and Disinfectants Maintenance
E.2-2
Disinfectants
E.2
You can use commercially available disinfectants with the given
restrictions for disinfecting the patient table and the accessories
as classified below:
E.2
Aldehyde
Aldehyde decompositors
Alcohols
Quaternary compounds
Organic acids
Peroxide compounds
B
.2
E
Use of disinfectants containing aldehyde or aldehyde decom-
positors will discolor the surfaces of therapy mat, positioning
mat, coronal prone head holder and CFK head holder.
E.2
Use of disinfectants containing alcohols or quaternary com-
pounds will impair the surface of immobilization straps.
E.2
i
.2
E
Organic acids and peroxide compounds can be used without
restriction.
E.2