syngo CT 2007E Emotion Operator Manual
02/2008
0.0
Daily Quality Measurements Quality Assurance
F.2-8
The following evaluations are shown:
F.2
ROI
Mean value of the CT value
Sigma value (measure for pixel noise)
F.2
The measurements are repeated automatically for each tube
voltage.
F.2
Stopping measurement
F.2
You can cancel quality measurement in any phase.
F.2
F.2
Click on the Cancel button.
The Quality/Daily measurement is aborted.
F.2
You have to repeat the quality measurement from the begin-
ning before you start the actual examinations.
F.2