Quality Assurance Constancy Test
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.3-7
0.0
Starting the constancy test
F.3
A reference measurement follows the same sequence as a nor-
mal constancy test.
F.3
F.3
✧ Confirm selection of the constancy test with Go.
You are prompted to enter the name of the tester and the serial
numbers of the phantoms.
F.3
F.3
✧ Enter your name and the data required.
F.3
✧ Click on Go.
✧ Prepare the measurement as described. (→ Page F.1-1
General Information)
F.3
✧ Click on Go when you have completed your preparations.
The “Quality Assurance Patient” is selected.
F.3
You are prompted to do press Start.
F.3
F.3
✧ Press the Start key on the control box.
The phantom position and the gantry tilt are checked.
F.3