syngo CT 2007E Emotion Operator Manual
02/2008
0.0
Constancy Test Quality Assurance
F.3-12
F.3
✧ Compare the position of the short / long strip on the image
with the options displayed in the content area of the Qual-
ity Constancy dialog box.
F.3
✧ Correct the table position by clicking Feed In or Feed Out.
i
.3
F
With Feed In and Feed Out, you can move the table by 1 mm.
F.3
✧ Record a new image and compare the position of the strips
again.
✧ Repeat image recording and correction of the table posi-
tion until you have set the correct z-position.
F.3
✧ Click on Continue.