Quality Assurance Constancy Test
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.3-17
0.0
Performing the homogenei-
ty test
F.3
With this test, you measure the homogeneity of the CT values in
5 regions of the water phantom.
F.3
F.3
✧ Press the Start key on the control box.
Scanning is started.
F.3
i
.3
F
After the previous test has been completed, the Homogeneity
test is started automatically.
F.3
An image of the water phantom is recorded. 5 ROIs are marked
in the image, one central and four peripheral.
F.3
In addition to the ROIs, the mean values and standard devia-
tions of the CT values are displayed.
F.3
The difference between the mean values of the central ROI and
the peripheral ROIs is displayed.
F.3
F.3