Quality Assurance Constancy Test
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.3-29
0.0
Exiting the constancy test
F.3
After the final test has been completed, you terminate the
monthly constancy test and return to the Home Menu dialog
window.
F.3
In the status and error message area you are prompted to
press Done.
F.3
F.3
Click on Done.
The Constancy test is terminated.
F.3
B
.3
F
In Reference mode, click on the Save button first before termi-
nating the constancy test with Done. Without Save the old ref-
erence values will be retained.
F.3
F.3
Click on Home.
You return to the Home window.
F.3
i
.3
F
You can now view the measurement report with the results of
the actual and previous tests. (→ Page F.3-30 Output of results)
F.3
Close the Local Service window.
You return to the syngo user interface.
F.3
Confirm closing of the service session with OK in the dialog
displayed.