Quality Assurance CTDI Test
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.4-9
0.0
Performing the measu rement
F.4
F.4
Click on Continue.
You are prompted to press Start.
F.4
F.4
Press the Start key on the control box.
F.4
Read off the value on the dosimeter and enter it in the pro-
tocol.
Continue with the next kV values.
Then insert the perspex rod with the measuring chamber
into position B of the phantom.
F.4
Click on Continue to start the second measurement.
i
.4
F
The second measurement might not involve all positions tested
in the first measurement.
F.4
When you have completed CTDI measurement on the CTDI
head phantom continue in the same way with the CTDI
body phantom.