Quality Assurance CTDI Test
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.4-9
0.0
Performing the measu rement
F.4
F.4
✧ Click on Continue.
You are prompted to press Start.
F.4
F.4
✧ Press the Start key on the control box.
F.4
✧ Read off the value on the dosimeter and enter it in the pro-
tocol.
✧ Continue with the next kV values.
✧ Then insert the perspex rod with the measuring chamber
into position B of the phantom.
F.4
✧ Click on Continue to start the second measurement.
i
.4
F
The second measurement might not involve all positions tested
in the first measurement.
F.4
✧ When you have completed CTDI measurement on the CTDI
head phantom continue in the same way with the CTDI
body phantom.