Quality Assurance Low Contrast Test (Optional)
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.5-11
0.0
Setting scan and recon param-
eters
F.5
You must set special scan and recon parameter to perform the
low contrast test adequately.
F.5
F.5
F.5
✧ Enter or select the required parameter in the subtask cards
Routine and Recon.
Scan parameter
F.5 F.5
mAs
F.5
130
F.5
kV
F.5
130
F.5
Scan time
F.5
1.0 s
F.5
Slice
F.5
10.0 mm
F.5
No. of scans
F.5
1
F.5
Recon parameter
F.5 F.5
Kernel
F.5
B20s
F.5
FoV
F.5
150 mm
F.5