Constancy Test of the Monitor Quality Assurance
F.7-4
Starting the monitor test
F.7
✧ Load the test image of your choice in the Viewing task card.
B
.7
F
To ensure that the criteria of the constancy test are met, make
sure that the requirements for the room lighting are fulfilled or
restored according to the acceptance test performed during ini-
tial start up.
F.7
✧ Perform the individual tests as described. (→ Page F.7-6
The individual tests)
✧ Enter the test results in the forms supplied. (→ System
Owner Manual)
i
.7
F
All measurement and test results must be confirmed with the
date and initials.
F.7
fB
.7
F
Test is out of tolerance.
F.7
✧ Contact Siemens Customer Service.