syngo CT 2007E Emotion Operator Manual
02/2008
0.0
O-13
0.0
Partial volume artifact
O.0
Artifact caused by bone structures which are only partially
included in the slice being scanned.
O.0
Phantom
O.0
Measuring object whose weakening properties is similar to
human; Reference body used for functional and quality assur-
ance tests.
O.0
Pitch factor
O.0
Term applicable only for spiral CT. It is defined as the ratio of
table feed per rotation to the collimated width of the entire
detector. Example: 16*0.75mm collimation, table feed 18 mm/
sec. This gives a volume pitch of 18 / 0.75 =24 and a pitch factor
of 18 / (16*0.75) = 18/12 = 1.5.
O.0
Pixel
O.0
Image element of the two dimensional image.
O.0
R
O.0
RAM
O.0
Fast, non-permanent memory (Random Access Memory) with
which the computer operates. The size of the available RAM
determines other factors, including the processing speed. Data
are stored long-term on hard disk, MOD or CD-R.
O.0
Raw data
O.0
Preprocessed measurement data. A raw data file corresponding
to each scan is used to reconstruct the CT image.
O.0
Real Time Display
O.0
Displaying of the measured image synchronous to the scanning
and just after the acquisition of a slice (rotation) is finished.
O.0