Sirona Dental Systems GmbH 2 Safety instructions
C3
+
2.3 On-site installation
59 06 404 D 3390
D 3390.201.01.18.02
11
2.3 On-site installation
The on-site installation must have been performed
according to our requirements. The details are described
in the document "Installation Requirements".
2.4 Installing the treatment center
Installation must be carried out by authorized techni-
cians in accordance with the installation instructions.
2.5 Media quality
The air and water supplies must meet the requirements
specified in the installation conditions.
To ensure compliance with the medical and national
legal requirements for water from treatment centers,
Sirona recommends equipping the treatment center with
a disinfection system. If you decide to operate the treat-
ment center without a disinfection system, you must
make alternative arrangements to ensure good water
quality.
As the owner of the treatment center, you are gener-
ally responsible for the water quality.
The colony count must fulfill the national regulations for
drinking water and must not exceed 500 CFU/ml under
any circumstances (CFU: colony forming unit).
In the event of an increased colony count, the indoor
installation must be checked and the cause of contami-
nation eliminated. Alternatively, a stand-alone water
supply can also be installed.
Before installation of the treatment center, an acceptable
microbiological water quality for the domestic water sup-
ply must be ensured and documented in the form of a
colony count. Sampling and the colony count must be
performed by a competent laboratory.
Check the treatment center water quality at regular inter-
vals; see "Microbiological inspection of the water" in the
document "Care and Cleaning by the Practice Team".
Please contact your specialized dealer or your relevant
dental association for the respective national require-
ments and measures. Where necessary, you must make
alternative arrangements to ensure good water quality if
you decide to operate the treatment center without a dis-
infection system.
Further information on the media quality requirements
are listed in the "Installation conditions" document.
Highly immunosuppressed patients should not come in
contact with water from the treatment center. The use of
sterile solutions is recommended.