Sirona Dental Systems GmbH 12 Assistant element
C3
+
12.5 Satelec Mini LED curing light
59 06 404 D 3390
D 3390.201.01.18.02
77
12.5 Satelec Mini LED curing light
Safety instructions
The curing light contains powerful LEDs. In accordance
with IEC 62471, they are classified in risk group 2 with
the light bar removed.
• Do not stare into the beam path for longer periods when
operating, as this can cause eye damage.
• Do not under any circumstances stare into the beam
path when the light bar is removed.
• Do not look into the light reflected by the tooth surface.
• Do not observe the light aperture with optical instru-
ments which can decrease the beam cross-section
(e.g. magnifying lenses).
• Use the glare shield!
• The light beams emitted by this instrument may be
hazardous and may never be directed to the eyes of any
person, even if this person wears goggles with protective
lenses.
• The light may be directed exclusively to the treatment
area in the oral cavity.
Heavy dips in the line voltage may lead to the Mini L.E.D.
being switched off. The fault is reported by a double
audible signal. Check the selected operating mode and
restart the Mini LED.