11 Basic treatment center settings in the Mode dialog Sirona Dental Systems GmbH
11.12 Instrument settings via the Mode dialog: REMOVE INSTRUMENT C3
+
59 06 404 D 3390
70 D 3390.201.01.18.02
11.12 Instrument settings via the Mode dialog:
REMOVE INSTRUMENT
This Mode dialog enables the following settings
• Instrument light voltage and
• Maximum torque of SL motors
Setting the instrument light voltage
You can change the voltage setting for instruments in
this Mode dialog.
First a message will appear prompting you to pick up an
instrument.
Then the current voltage setting of this instrument is dis-
played.
You can adjust the instrument light voltage in increments
of 0.1 V from 3.0 to 5.6 V with the
– / + keys.
WARNING is displayed when the voltage exceeds 3.9 V.
After placing back the instrument, pick up and set the
next instrument.
Observe the voltage specifications provided by the
instrument manufacturer!
For Sirona instruments, the voltage (3.6 V) is already
preset.
Setting the torque of the electric motors
This Mode dialog is used to set the automatic controller
action of the SL or SL ISO electric motors. Individual set-
tings are possible for each motor holder.
First a message will appear prompting you to pick up an
instrument.
After removing the instrument, select the setting dialog
by pressing the Mode key again:
MAX. TORQUE
If Yes is flashing-
The motor will attain a higher torque at speeds
> 15,000 rpm.
If NO is flashing-
The running smoothness of the motor will be higher at
speeds > 15,000 rpm.