8 Positioning patients Sirona Dental Systems GmbH
8.6 Programming patient positions C3
+
59 06 404 D 3390
46 D 3390.201.01.18.02
8.6 Programming patient positions
The four factory-set programs can be changed indi-
vidually by user A and user B.
•Programs 1 and 2
Entry / exit program 0
Mouth rinsing program S
1. Move the chair to the desired treatment position by
pressing the different setting keys (see page 45).
2. To make the operating light switch on or remain off
when the chair reaches the programmed treatment
position, you must switch the lamp ON or OFF now.
This setting is then also stored in the program.
NOTICE
In order to avoid damage to the dentist and assistant
elements and to the chair upholstery, make sure that
these elements do not protrude into the movement range
of the chair.
3. To save the program settings, press the desired pro-
gram key for approx. 3 seconds until an audible sig-
nal sounds and the LED of the corresponding key
lights up.
Programming is now complete.
IMPORTANT
Programming is not possible by activating the program
keys with the foot switch (cursor control). Thus pro-
gramming errors are avoided.