2 Safety instructions Sirona Dental Systems GmbH
2.12 Intermittent operation C3
+
59 06 404 D 3390
14 D 3390.201.01.18.02
Do not spray liquids such as disinfectants into the venti-
lation slots. This may lead to malfunctions. Use only
wipe disinfection in the vicinity of the ventilation slots.
2.12 Intermittent operation
Intermittent operation
The motors of the treatment center and of the treatment
instruments are designed for intermittent operation cor-
responding to the dental mode of treatment.
Drive motors for patient chair and backrest: max. 6%
duty cycle (15 s "ON"/3 min, 55 s "OFF")
Other motors: Intermittent operation, max. 10% duty
cycle (max. 2 min "ON"/18 min "OFF")
2.13 Care, cleaning agents, and disinfectants
Unsuitable care and cleaning agents and disinfectants may
corrode the surface of the unit or impair its functioning.
Therefore, use only care and cleaning agents and disin-
fectants which have been approved by Sirona. For fur-
ther information, see the document "Care, cleaning
agents, and disinfectants", REF 59 70 905.
2.14 Modifications and extensions of the system
Modificat ions and extensio ns of the system
Modifications to this system which might affect the safety
of the system owner, patients or other persons are pro-
hibited by law.
For reasons of product safety, this product may be oper-
ated only with original Sirona accessories or third-party
accessories expressly approved by Sirona. The user is
responsible for any damage resulting from the use of
non-approved accessories.
If any devices not approved by Sirona are connected,
they must comply with the applicable standards, e.g.:
IEC 60950-1 for information technology equipment
(e.g. PC) and
IEC 60601-1 for medical electrical equipment.
The treatment center monitor must fulfill the require-
ments of the IEC 60950-1 standard.
The loudspeaker port of the monitor may only be con-
nected to a device that complies with IEC 60950-1
(e.g. a PC) or IEC 60601-1. Under no circumstances
may it be connected to a stereo system or similar.