3.2 Technical data
Dental treatment center
Model designation: C3
+
Power connection: 100 V / 115 V / 230 VAC 50 / 60 Hz
Rated current: 11.5 A at 100 VAC
9.5 A at 115 VAC
4.5 A at 230 VAC.
In addition, max. 6 A for external devices
Overvoltage category: 2 acc. to IEC 60664-1
Main unit fuse: for 100/115 V AC: T 10 A H, 250 V AC , REF 10 77
460
for 230 V AC: T 6.3 A H, 250 V AC , REF 10 77 452
Protection class: Class I device
Device class in accordance
with Directive 93/42/EEC:
Class IIa equipment
Degree of protection against elec-
trical shock:
Type B applied parts
Except SIROCAM 3 / SIROCAM C. These are:
Type BF applied parts
Degree of protection against
ingress of water:
Ordinary equipment (without protection against
ingress of water)
The foot switch has an IP X1 degree of protection
against liquids (drip-proof).
Mode of operation Continuous operation with intermittent loading accord-
ing to the type of dental operation, see "Intermittent
operation" on page 14
This device is permanently installed. Operation is not
permitted in mobile vehicles.
Transport and storage conditions: Temperature: -40°C – +70°C
-40°F – +158°F
relative humidity: 10% – 95%
Air pressure: 500hPa – 1,060hPa
Operating conditions: Ambient temperature: 10°C – 40°C
50°F – 104°F
relative humidity: 30% – 85% (no condensa-
tion)
Air pressure: 700hPa – 1,060hPa
Installation location: ≤ 3,000 m above sea level
This treatment center is not suitable for operation in
areas subject to explosion hazards.
Pollution degree: 2 acc. to IEC 60664-1
Tests/Approvals: See "Standards / Approvals" on page 17
Year of manufacture: (on the rating plate of the patient chair)