EasyManua.ls Logo

STONEX R25LR - Horizontal Collimation Error C

STONEX R25LR
128 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Stonex R25/R25LR Total Station - User Manual 110

 
When the offset value falls within the range ±20″, adjustment is not necessary. If one of the offset values (Xoffset,
Yoffset) exceeds ±20″, adjust the value as follows.
II. Adjustment by program
In the second page of the menu select F1[ADJ],
Then, F4 “Tilt Offset” to adjust the tilt.
Press F3[ENT] to set the horizontal angle to 0°.
“Take F2is displayed. Rotate the top of instrument 180°/200gon. Wait
a few seconds for the display to stabilize, then press F3[ENT].
If both correction constant are in the range, press F3[YES] to renew
the correction angle, or press F4[NO] to give up this adjustment,
10.8. Horizontal collimation error C
If the telescope’s sight line isn’t perpendicular to the horizontal axis, the collimation error will appear. The
assembling, transportation and operation will cause this error.
I. Check

Table of Contents

Related product manuals