# Error message Explanation Action
20 Question
Are you sure you want
to delete all preparation
groups and methods?
Hint
You
cannot
undo
deletion
of
groups
and
method
s.
Press Enter to delete all
the groups and methods.
Struers methods are not
deleted.
101 Warning
Cannot keep preset
force.
The machine could not
maintain the selected force.
Make sure that the air
pressure is 6 bar. Re-
start the process.
If the error remains,
contact Struers Service.
114 Warning
Disc motor:
Warning: General.
Warning code: 0
Non-specific warning reported
by movement unit.
Re-start.
If the error remains,
contact Struers Service.
Make a note of the
reason code displayed.
124 Warning
Water for polishing
valve, bad electrical
connection detected.
Bad wiring in the valve
controlling polishing water.
Re-start.
If the error remains,
contact Struers Service.
125 Warning
OP suspension flushing
valve, bad electrical
connection detected.
Bad wiring in the valve
controlling OP flushing.
Re-start.
If the error remains,
contact Struers Service.
126 Warning
Disc cooling valve, bad
electrical connection
detected.
Bad wiring in the valve
controlling disc cooling.
Re-start.
If the error remains,
contact Struers Service.
141 Warning
No air connected or air
pressure too low.
The air supply is not
connected or the pressure is
too low.
Check the air supply
connection.
Make sure that the air
pressure is min. 6 bar.
50 AbraPol-30
8 Troubleshooting - AbraPol-30