Installing the ESU Driver Upgrade
101 EN
Caution Programming the digital board may take several
minutes. To avoid equipment damage, please do not
remove the dongle card before Sidne™ says, “Please
remove card.”
6. Wait as Sidne™ restarts. When the console says, “Sidne™
ready,” the driver is installed and ready for use.
Caution Sidne™ should say “ESU Connected” and the EDAM
LED should be flashing green within 20 seconds of
powering On the ESU. A different LED output may
indicate device problems. If the LED does not flash
green after 20 seconds, refer to the “Troubleshooting
Guide” of this addendum for instructions.
7. Power On the ESU.
Please retain this addendum for future reference.
When Sidne™ says this... do this...
“Software card detected. Please stand
by.
Wait for the next instruction.
“Programming 10% complete.
*
* Sidne™ will announce the completion percentage
throughout the installation process. Do not remove the
dongle until Sidne says, “please remove card.”
Wait for the next instruction.
“Programming failed.
**
** Sidne™ will announce this message only if the
programming has failed.
Remove and reinsert the dongle.
“Programming complete. Wait for the next instruction.
“Please remove card. Remove the driver dongle from the
Sidne™ console.