SIDNE™ Cleaning and Maintenance
63 EN
Cleaning
Warning Unplug the Sidne™ console from the electrical outlet
before cleaning the unit.
Caution Do not immerse the console, tablet, or headset in any
liquid, as product damage will result.
Caution Do not use solvents, such as alcohol, or cleaning
solutions that contain ammonia to clean the console,
tablet, or headset, as product damage may result.
Caution Do not sterilize the console, tablet, or headset, as
product damage may result.
1. Wipe the console with a soft cloth dampened in a mild cleaning
solution.
2. Clean the console with disinfectant if needed.
3. Perform the same two steps for cleaning the tablet and headset.
Maintenance
The Sidne™ system requires no preventative or periodic maintenance.
The fuse on the console may be replaced at the hospital by a qualified
technician.
Warning To reduce the risk of electrical shock, do not open the
Sidne™ console, tablet, or headset. There are no user-
serviceable components inside. Should service be
needed, notify your local Stryker representative.
Replacing Fuses
1. Unplug the Sidne™ console from the electrical outlet and
disconnect the power cord from the rear console panel.
2. Unlatch the fuse holder next to the AC connection on the rear
console panel and remove the fuse. You may need to press the
tab on the fuse holder with a slender screwdriver to release the
latch.
3. Replace the fuse with a fuse of the same value and rating,
pushing gently until the tab snaps in place.
Warning To avoid the risk of fire, use only replacement fuses
specified in the Specifications section of this manual.
SIDNE™ Cleaning and Maintenance