92 EN
This section is intended to give an outline of the start-up instructions for
experienced users. Refer to the relevant section(s) of this addendum for
further instructions.
1. Ensure that the ESU and Sidne™ are powered Off.
2. Install the cover plate (if necessary).
3. Connect the ESU to Sidne™.
a. Connect the EDAM adapter cable to the ESU.
b. Connect the EDAM adapter cable to the EDAM.
c. Connect the Sidne™ interface cable to the EDAM.
d. Connect the Sidne™ interface cable to Sidne™.
e. Mount the EDAM.
4. Power On the Sidne™ console.
5. Install the ESU driver upgrade.
6. Sidne™ will restart. Wait for Sidne™ to say “Sidne™ ready”
before powering On the ESU.
Caution Sidne™ should say “ESU Connected” and the EDAM
LED should be flashing green within 20 seconds of
powering On the ESU. A different LED output may
indicate device problems. If the LED does not flash
green after 20 seconds, refer to the “Troubleshooting
Guide” of this addendum for instructions.
7. Power On the ESU.
This completes the installation procedure.
Quick Start-Up Guide