72 EN
Testing Compatibility
1. Identify any critical care medical devices located within the
communication range of Sidne™.
2. Test each device for compatibility with the Sidne™ system.
Power on the critical care devices.
Perform a variety of functions with the Sidne system. Issue
commands with both the tablet and headset, and power on and
off the tablet, headset, and console.
Observe the critical care devices for per function.
Note Should a device demonstrate irregular behavior while
Sidne is in operation, increase the distance between the
device and the Sidne console and repeat step 2 until
normal behavior is restored. Consult the tables below for
guidance in placing the Sidne console.
Note For extra information about this test method, please refer to
the “Electromagnetic Compatibility” section in this manual,
or to the ANSI C63.18 standard.
Note The radio-frequency communication range of Sidne™ has
been moderated to 30-40 feet to help manage
electromagnetic interference.