Troubleshooting Guide
107 EN
Table 4 shows the most common errors and possible solutions to each.
Troubleshooting Guide
Problem Possible Solution
1. No access to
ESU through
Sidne™
This problem has several possible causes.
Please see solutions for each problem:
a. EDAM LED
stays flashing
red
*
Contact Stryker Endoscopy or your local
Stryker sales representative for firmware
upgrade.
b. EDAM LED
stays solid
red*
The ESU is not communicating properly with
the Sidne™ ESU EDAM. Ensure that the
ESU is properly connected to Sidne™. See
the “Connecting Sidne™ to the Valleylab
Force FX™ Electrosurgical Unit” section of
this addendum for more details. Attempt the
following solutions in order:
1. Power Off ESU.
2. Ensure that the ESU EDAM adapter
cable is plugged into the EDAM.
3. Ensure that the ESU EDAM adapter
cable is plugged into the ESU.
4. Power On ESU.