108 EN
c. EDAM LED
stays solid
yellow*
Sidne™ is not communicating properly with
the Sidne™ ESU EDAM. See the
“Connecting Sidne™ to the Valleylab Force
FX™ Electrosurgical Unit” section of this
addendum for more details. Attempt the
following fixes in order:
1. Power Off ESU.
2. Ensure Sidne™ is powered On.
3. Ensure that the Sidne™ interface cable
is plugged into the EDAM.
4. Ensure that the Sidne™ interface cable
is plugged into Sidne™.
5. Power On ESU.
2. Sidne™ says
“ESU
Disconnected”
Power Off ESU.
Ensure that the ESU is properly connected to
Sidne™. See the “Connecting Sidne™ to the
Valleylab Force FX™ Electrosurgical Unit”
section of this addendum for further
instructions.
Power On ESU.
3. Sidne™ says
“Check patient
return
electrode”
Refer to the Valleylab Force FX™
Electrosurgical Unit Manual for further
details.
4. Sidne™ says
“Error. Check
generator”
Refer to the Valleylab Force FX™
Electrosurgical Unit Manual for further
details.
* The EDAM LED should flash green when the system is functioning
normally.
Problem Possible Solution