Electromagnetic Compatibility Tables
109 EN
Like other electrical medical accesories, the Sidne ESU EDAM requires
special precautions to ensure electromagnetic compatibility with other
electrical medical devices. To ensure electromagnetic compatibility
(EMC), the Sidne ESU EDAM must be installed and operated according
to the EMC information provided in this manual.
Warning Do not use cables or accessories other than those
provided with the Sidne ESU EDAM, as this may result
in increased electromagnetic emissions or decreased
immunity to such emissions.
Warning If the Sidne ESU EDAM is used adjacent to or stacked
with other equipment, observe and verify normal
operation of the Sidne ESU EDAM in the configuration
it will be used in prior to using it in a surgical
procedure. Consult the tables below for guidance in
positioning the Sidne ESU EDAM.
Caution Portable and mobile RF communications equipment
may affect the normal function of the Sidne ESU EDAM.
The Sidne ESU EDAM has been designed and tested to comply with IEC
60601-1-2:2001 requirements for EMC with other devices.