Guidance and Manufacturer's Declaration: Electromagnetic Immunity
Sidne ESU EDAM is intended for use in the electromagnetic environment specified below. The
customer or the user of Sidne ESU EDAM should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment:
Guidance
Electrostatic
Discharge (ESD)
IEC61000-4-2
±6kV contact
±8kV air
±2,4,6kV
contact
±2,4,8kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
±2kV for power
supply lines
±1kV for input/
output lines
±2kV for
power supply
lines (Sidne)
±1kV for
input/output
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power-frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE: Ut is the a.c. mains voltage prior to application of the test level.
Portable and mobile RF
communications equipment should be
used no closer to any part of the Sidne
ESU EDAM system, including its
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz
3 V
Recommended Separation Distance
d1.17P=