106 EN
Sidne™ Feedback Messages
Table 3 shows the Sidne™ feedback messages and their meanings.
Sidne™ Feedback Messages
Sidne™ Voice Message Meaning
“Connected” ESU is turned On and plugged into Sidne™.
“Disconnected” ESU has been disconnected or turned Off
“Check Patient Return
Electrode”
Return Electrode Monitor fault has been detected. “video Error” is
displayed above the status bar for 10 seconds.
“Error. Check Generator” ESU generator error code has been detected. Check ESU panel
display. All commands are grayed out and “video Error” is displayed
until the ESU is turned Off and On.
“ESU... Added as a prefix to the above messages, when a feedback message
is announced while not in the ESU Command Mode. Also added as a
prefix to the “Connected” and “Disconnected” messages.