Setting Up the Sidne™ System
19 EN
Connect video inputs, such as cameras, to an XGA port (see
Figure 11), or a S-Video port (see Figure 10).
To take advantage of the high definition support provided by
Sidne™, connect the DVI/XGA inputs to a high definition
camera system such as the Stryker 1088 Medical Video
Camera. Ensure that the camera is set to high definition mode.
5. Connect any video outputs.
Connect any video outputs, such as monitors or recording
devices, to the S-video-out port (see
Figure 10).
Note When connecting video and audio inputs and outputs,
make Sidne™ the first connection in the loop. For example,
rather than connect a camera directly to a monitor or
recording device, connect it directly to Sidne™. After
connecting the camera to Sidne™, connect the monitor
and recording device to Sidne™.
6. Connect Sidne™ to a telephone line if desired (see Figure 9).
Note The telephone option is only available if the supplemental
telephone driver has been installed.
Connect a telephone cord from a telephone wall jack to the
line-in jack on the rear Sidne™ console.
Connect any external telephone to the phone jack on the rear
Sidne™ console.
7. Connect the tablet (see Setting Up the Tablet) and the headset
(see Setting Up the Headset) (see also
Figure 9).
8. Apply power to the devices connected to the Sidne™ console.
9. Apply power to the Sidne™ console.
For first use only: Switch the on/off switch on the rear console
panel to on (represented by the “I” symbol).
Note Turning on the unit using the on/off switch on the rear
console panel will fully power up the unit. In this case, it is
not necessary to press the on/off button on the front
console panel.
For every subsequent use: Press the on/off button on the front
console panel to activate the system.
After the Sidne™ system boots up, the console says,
“Greetings. One moment please. Sidne™ ready.” The Sidne™
system will review the connected operating-room devices, list
them in order, and announce any device error conditions. For
example: “Insufflator/ Camera/ VCR connected.”