XLi Time & Frequency System 145
XLi-man, Issue 8, 6/17/2008, Rev. H
2
5
1
SSSSSSSSSSSS SSSSSSSSSS SSSSSSS S SSS S SSSSS S
wildcard-digit in the time field is used to specify the Start (or Stop) Time of the repetitive pulses, which in
turn specifies the pulse width of the repetitive pulses. The least significant “wildcard” character (the one
to the immediate left of the most significant non-wildcard-digit) specifies the period of repetition.
When specifying repetitive rates, the Start Time must include the same number of significant digits as the
Stop Time or an ambiguous output may occur.
Any time the clock reads a new time that matches the specified least significant digits, a pulse either
starts or stops. For example, the following string produces a one-second pulse at midnight on every day
of the year with a “1” in the least significant digit:
F111 PPO XX1:00:00:00.000000 XX1:00:00:01.000000<CR>
For example, to produce a repetitive 5 microsecond pulse occurring every 10 microseconds (i.e.,
repetitive pulses with 100 kHz frequency with the start time or rising-edge on-time and the stop time or
falling-edge at 5
μS- or 10 μS pulse period with 5 μS pulse width), enter:
F111 PPO XXX:XX:XX:XX.XXXXX0 XXX:XX:XX:XX.XXXXX5<CR>
The XLi responds:
OK<CR><LF>
Or, for example, to produce a repetitive 50-microsecond pulse occurring every 100 microseconds (i.e.,
repetitive pulses with 10 kHz frequency with the start time or rising-edge at 5
μS and the stop time or
falling-edge at 55
μS - or 100 μS pulse period with 50 μS pulse width), enter:
F111 PPO XXX:XX:XX:XX.XXXX05 XXX:XX:XX:XX.XXXX55<CR>
The XLi responds:
OK<CR><LF>
Or, for example, to produce a repetitive 50-microsecond pulse occurring every 100 milliseconds (i.e.,
repetitive pulses with 10 PPS frequency with the start time or rising-edge at 5
μS and the stop time or
falling-edge at 55
μS - or 100ms pulse period with 50 μS pulse width), enter:
F111 PPO XXX:XX:XX:XX.X00005 XXX:XX:XX:XX.X00055<CR>
Clock Reading
Pulse
001:00:00:00.000000 Start
001:00:00:00.000001 No Change
001:00:00:01.000000 Stop
001:01:00:00.000000 No Change
001:00:00:00.000000 Start (etc...)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Symmetricom XLi and is the answer not in the manual?

Symmetricom XLi Specifications

General IconGeneral
BrandSymmetricom
ModelXLi
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals