XLi Time & Frequency System 163
XLi-man, Issue 8, 6/17/2008, Rev. H
2
5
1
SSSSSSSSSSSS SSSSSSSSSS SSSSSSS S SSS S SSSSS S
To change the INPUT MODE and HQ TIME REFERENCE, send a command using the following format:
F123<SP>B<N><SP>C<SP><CONFIG><SP><REF><LT>
where:
For example, enter:
F123 B3 C SEC HAVE QUICK 1PPS
The XLi changes INPUT MODE to HAVE QUICK 1PPS, changes HQ TIME REFERENCE to
SECONDARY, and responds:
OK
F = ASCII character F.
123 = function number.
<SP> = ASCII space character.
B = ASCII character B to denote Option Bay number follows
<N> = Option Bay Number, 1 through 10.
C = ASCII character to denote “Configuration of Time Reference”
<CONFIG> = Option Board Configuration: PRI, SEC, or STBY
<REF> = Option Board Reference: 1PPS, HAVE QUICK, or HAVE QUICK 1PPS
<LT> = line terminator; for output strings (a carriage return and line feed) for input strings
(a carriage return only).