68 XLi Time & Frequency System
XLi-man, Issue 8, 6/17/2008, Rev. H
SSSSSSSSSSSS SSSSSSSSSS SSSSSSS S SSS S SSSSS S
1
F27 B<N> PS -1.00<cr><lf>
Response:
OK<cr><lf>
Request current Time Deviation Preset Value. Enter:
F27 B<N> PS <cr><lf>
Response:
F27 B<N> PS= -1.00<cr><lf>
Preset accumulated Time Deviation to 0.0 seconds. Enter:
F27 B<N> PS 0<cr><lf>
Response:
OK<cr><lf>
Preset accumulated Time Deviation to 100.0 seconds. Enter:
F27 B<N> PS 100<cr><lf>
Response:
ERROR 01 VALUE OUT OF RANGE<cr><lf>
F27 B<N> TD<cr><lf> DATA ON DEMAND OUTPUT
The Data On Demand Output allows the user to request a single measurement by sending an upper
case "T<cr><lf>" while in the "F27 B<N> TD" mode. Subsequent receipts of the letter T (without the
carriage return linefeed combination) are responded to with the current data values.
Local Time of receipt of the 'T' to the millisecond is recorded, System Time is calculated from the
recorded Local Time, and the complete FTM data string is transmitted back to the user. An example
session follows:
F27 B<N> TD<cr><lf>
T
069:15:25:27.545QT+00.477F-0.011SF+59.989ST15:25:28.022
T
069:15:25:31.932QT+00.477F-0.013SF+59.987ST15:25:32.409
T
069:15:25:32.524QT+00.476F-0.012SF+59.988ST15:25:33.000
^C
OK<cr><lf>
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com