EasyManuals Logo

Symmetricom XLi User Manual

Symmetricom XLi
318 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
XLi Time & Frequency System 79
XLi-man, Issue 8, 6/17/2008, Rev. H
2
5
1
SSSSSSSSSSSS SSSSSSSSSS SSSSSSS S SSS S SSSSS S
To calculate what the setting should be, multiply the delay/foot by the length of the cable in feet. The
typical delays for the following cable types are:
• RG-58 – approximately 1.4 ns/foot
• RG-59 – approximately 1.24 ns/foot
Command Line
To display the current distribution cable delay, enter:
F52<CR>
The XLi responds using the following format:
F52<SEP><SIGN><DELAY>ns<CR><LF>
where:
For example, to display the current distribution cable delay, enter:
F52<CR>
XLi responds:
F52 +000000ns<CR><LF>
To set the distribution cable delay to 60 nS, enter:
F52 +000060nS<CR>
XLi responds:
OK<CR><LF>
F = ASCII character F (f or F for input string).
52 = the function number.
<S> = one or more space characters.
<SIGN> = either + or –
<D> = 1 to 6 digit delay from +999999 nS to –999999 nS
ns = nanoseconds (ns or NS for input string)
<CR> = carriage return character
<LF> = line feed character
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Symmetricom XLi and is the answer not in the manual?

Symmetricom XLi Specifications

General IconGeneral
BrandSymmetricom
ModelXLi
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals