EasyManua.ls Logo

Teac TD-9000T - Page 34

Teac TD-9000T
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
34
4. Settings
4-3-1-6. LINEARIZE CALIB.
The linearity of load measurements is improved by increasing
the calibration points.
After equivalent input calibration, input the linearization
calibration.
1 Touch the ON button to enable linearization
calibration.
2 Touch the number of the calibration point to input
on the right side of the screen.
3 Input the calibration value and touch OK.
The “LOAD VAL. cannot be set to a value that would be
5% or more relative to the straight-line (reference curve)
that connects the no load and rated load outputs.
ATTENTION
The following conditions must always be met.
o 0 < calibration value < maximum rated output
o Value of lower-numbered calibration point
< value of higher-numbered calibration point
4 Conrm the input value. Then, touch ESC to return
to the previous screen.
5 Repeat steps 2–4 to input necessary calibration
values, and touch the OK button.
NOTE
o The calibration waveform shown on the LINEARIZE
(EQUIVALENT CALIB.) screen is an illustration, not the actual
calibration curve.
o After linearization calibration value input, touch the OFF
button to disable linearization calibration. Since input cal-
ibration values are saved, touching the ON button will
enable linearization calibration.
o To disable a specific calibration point, touch its number so
that its background color disappears
In the illustration below, calibration point 2 is disabled.
o The possible range of linearization is shown by the dashed
lines in the illustration below.
Rated
output
Rated
capacity
Rated output + 5%
Rated output − 5%
Output
Capacity
0
Reference
curve

Table of Contents

Related product manuals